WM 68401 04/2021
MEDUCORE Standard
2
EN 331
13 Operator menu
13.8.2 Possible values
Parameter
Possible
values
Description
Factory
setting
Initial cuff
pressure
Adult
120 mmHg -
280 mmHg,
in increments of 5
Here you can set the NIBP
cuff pressure to which the
device must inflate the NIBP
cuff for a particular patient
group.
180 mmHg
Child
80 mmHg -
170 mmHg,
in increments of 5
120 mmHg
Infant
60 mmHg -
140 mmHg,
in increments of 5
90 mmHg
Interval -
Off
0:30 min
0:45 min
1:00 min
1:15 min
1:30 min
1:45 min
2:00 min
3:00 min
5:00 min
10:00 min
15:00 min
30:00 min
60:00 min
Here you can set after how
long another NIBP
measurement is to be started
automatically.
Off
Venous stasis
pressure
-
60 mmHg -
140 mmHg,
in increments of 5
Here you can set the pressure
to which the NIBP cuff is
inflated for venous stasis
(IV function button).
80 mmHg
Venous stasis
duration
- 10 s - 120 s
Here you can set for venous
stasis (IV function button)
the time for which pressure is
to be maintained in the NIBP
cuff.
60 s