334 EN MEDUCORE Standard
2
WM 68401 04/2021
13 Operator menu
13.9.2 Possible values
Parameter
Possible
values
Description
Factory
setting
Session report -
Resuscitation
report
Vital signs
Events
Events and vital
signs
Here you can set the details
the session report contains.
Resuscitation
report
Vital signs
Events
Automatic stop
print
- 5 s to 120 s
Here you can set the time
after which printing of a live
printout or replay printout
finishes automatically.
10 s
Feed rate -
12.5 mm/s
25 mm/s
50 mm/s
Here you can preset the feed
rate at which the printer
prints a live printout and a
replay printout.
25 mm/s
Defibrillation
report
-
Activated
Deactivated
Here you can set whether a
defibrillation report is printed
in the event of defibrillation
or cardioversion (only with
the Cardioversion option) if
the printer is connected to
the device and is ready for
use.
Activated
Pair printer - -
Here you can pair the printer
with a device (see “12.2 Pair
printer with device (only with
Printing option)”, page 290).
The connection is always set
up automatically whenever
the device and printer are
switched on and are within
range.
-