36 EN MEDUCORE Standard
2
WM 68401 04/2021
3 Description
No. Designation Description
1 Line power indicator
Indicates that the device is being supplied by line
power.
2 Battery status indicator
• Green light on: The battery is full or is not being
charged because it is outside charging
temperature range.
• Green light flashing: The battery is being charged.
• Red light on: The battery is defective or not in the
device.
• No light: The device is not connected to line
power.
3 Shock button
Triggers an electric shock for defibrillation/
cardioversion.
4 Shock standby indicator
Red light flashes when the device is ready to deliver a
shock.
5 Alarm button
• Pauses audio alarm output for a certain length of
time.
• Mutes audio alarm output.
• Cancels audio alarm output.
• Deactivates muting of audio alarm output and
alarm cancellation.
6 Menu button
• In the start menu: Provides access to the operator
menu.
• In a mode: Provides access to the user menu.
• In the button function test during the function
check: Cancels the button function test.
7 Navigation knob
• Allows values to be selected (turn).
• Confirms selected values (press).
• In a mode: Provides access to the application
menu (press).
8 On/Off button
• With device switched off: Press briefly to switch
the device on.
• With device switched on: Press and hold to switch
the device off.
9 Function buttons
• Provide access to the mode shown in the display.
• Activate/deactivate the functions shown in the
display.
10 NIBP button
• Activates NIBP function mode (press NIBP button
< 2 s).
• Starts an NIBP measurement (press NIBP button
for > 2 s).