18 Technical data
The performance of the rhythm detector was validated with a
representative cross-section of ECG data from the following
databases:
• Creighton University Ventricular Tachyarrhythmia Database
• American Heart Association ECG Database
• MIT-BIH Malignant Ventricular Arrhythmia Database
Effectiveness results are shown in the tables below:
Rhythms Scope of sample
Objectives for
effectiveness
Sensitivity in % Specificity in %
Shockable
Coarse VF
(Amplitude
> 0.35 mV)
944 > 90 % 98.31 % -
Rapid VT
(f > 180/min)
252 > 75 % 94.05 % -
Non-shockable
Non-shockable
rhythms
3070 > 99 % - 99.64 %