WM 68401 04/2021
MEDUCORE Standard
2
EN 383
18 Technical data
Risk of injury from overly powerful high-frequency
interference signals or electromagnetic fields!
Overly strong high-frequency interference signals from portable
high-frequency communications equipment (e.g. radios) or
electromagnetic fields may lead to incorrect analysis, incorrect
measurement results, and incorrect alarms, and thus impair the
functioning of the device and injure the patient.
Maintain separation distances.
With portable high-frequency communications devices,
maintain a minimum distance of 30 cm from the device,
components, and accessories.
Recommended separation distances between portable and mobile HF
telecommunications devices and MEDUCORE Standard
2
MEDUCORE Standard
2
is intended for use in an electromagnetic environment in which HF interference
variables are controlled. The customer or user of the MEDUCORE Standard
2
can avoid electromagnetic
interference by maintaining a minimum distance between portable and mobile HF telecommunications
equipment (transmitters) and MEDUCORE Standard
2
(depending on the output power of the HF
telecommunications equipment, see below).
Rated power of
HF device in W
Separation distance depending on transmission frequency in m
150 kHz -
80 MHz
150 kHz -
80 MHz in the
ISM bands
When used as a monitor
80 MHz -
800 MHz
800 MHz -
2.5 GHz
0.01 0.12 0.12 0.4 0.77
0.1 0.38 0.38 1.3 2.4
1 1.2 1.2 4 7.7
10 3.8 3.8 13 24
100 12 12 40 77
Recommended separation distances between portable and mobile HF
telecommunications devices and MEDUCORE Standard
2
Rated power of
HF device in W
Separation distance depending on transmission frequency in m
When used as a defibrillator
Defibrillator: No unintended
energy delivery
150 kHz -
80 MHz
150 kHz -
80 MHz in the
ISM bands
80 MHz -
800 MHz
800 MHz -
2.5 GHz
0.01 0.12 0.27 0.06 0.12
0.1 0.38 0.66 0.15 0.38
1 1.2 2.7 0.6 1.2
10 3.8 6.6 1.5 3.8
100 12 27 6 12