EasyManuals Logo

Weinmann MEDUCORE Standard2 User Manual

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #385 background imageLoading...
Page #385 background image
WM 68401 04/2021
MEDUCORE Standard
2
EN 385
18 Technical data
Mobile HF communications equipment
Conducted interference induced
by high-frequency fields acc. to
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in the ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V in the ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips, short interruptions
and voltage fluctuations acc. to
IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase at 0°
0 % U
T
; 250/300 cycles
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase at 0°
0 % U
T
; 250/300 cycles
Electromagnetic immunity
Mobile HF communications equipment
Test
frequency
in MHz
Frequency
band in
MHz
a)
Radio
service
a)
Modulation
a)
Maximum
power
output
in W
Distance
in m
Interfer-
ence
immunity
test level
in V/m
385 380 to 390 TETRA 400
Pulse
modulation
b)
18 Hz
1.8 0.3 27
450 430 to 470
GMRS 460
FRS460
FM
c)
± 5 kHz
deviation
1 kHz sine
2 0.3 28
810
800 to 960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse
modulation
b)
18 Hz
2 0.3 28
870
930
1720
1700 to 1900
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3,
4, 25
UMTS
Pulse
modulation
b)
217 Hz
2 0.3 28
1845
1970

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals