Chapter 3 PERFORMING MEASUREMENTS
3-42
Changing the Splice Loss Threshold
When auto search is executed, splice losses exceeding the threshold set here will be
detected as events.
The splice loss threshold can be changed as explained below.
Example: “0.28dB” “0.35dB”
For details on splice loss, refer to page 1-41.
1. Locate the cursor to “SPLICE LOSS” by using the rotary knob or the arrow key.
2. Press [ENTER].
A selection window will appear.
Default setting: 0.03dB
3. Locate the cursor to “2” by using [W] / [X].
4. Change to “3” by using the rotary knob or [S].
TIP
Selectable
splice loss threshold
0.01dB
to
9.99dB
(0.01dB step)