Chapter 5 EDITING AUTOMATIC SEARCH RESULTS
5-15
5.3 Editing the Event List
Distance calculation is performed using the group index set in the measurement
conditions. However, if two or more optical fibers are connected as a result of installation
of new transmission system, the group index varies with the optical fibers. In this case,
distance calculation can be performed with higher accuracy by setting the group index of
each optical fiber. To obtain accurate analysis results, it is necessary to edit the value
between certain points.
This section explains the following event list edit functions.
• Editing an interval distance
• Editing a return loss
• Editing an interval group index