15.6 General Specifications
Item Specifications
Display
256 × 64 dots fluorescent tube (VFD)
Internal memory GS200ROM: 4 MB (non-volatile; used to save setup files and program files)
GS200RAM: 16 MB (volatile, cleared when the power is turned off; used to save the measured
results)
Warm-up time
Approx. 60 minutes
Storage conditions Temperature –15°C to 60°C
Humidity 20% RH to 80% RH (no condensation)
Elevation 2000 m or less
Operating conditions Temperature 5°C to 40°C
Humidity 20% RH to 80% RH (no condensation)
Elevation 2000 m or less
Rated supply voltage 100 VAC, 120 VAC, and 230 VAC
Permitted supply voltage Each rated voltage ± 10%
range
Rated power supply 50/60 Hz
frequency
Permitted power supply 48 Hz to 63 Hz
frequency range
Maximum allowable input
Across OUTPUT Hi and OUTPUT Lo and across SENSE Hi and SENSE Lo ±32 Vpeak
voltage Across OUTPUT Lo and G terminal and across SENSE Lo and G terminal
±42 Vpeak
Across each OUTPUT terminal and each SENSE terminal ±0.5 Vpeak
Across G terminal and case ±250 Vpeak
Maximum power Approx. 80 VA
consumption
Maximum common-mode Across each terminal and case
±250 Vpeak
voltage
External dimensions Approx. 213 (W) × 88 (H) × 379 (D) mm (excluding protrusions)
W
eight Approx. 5 kg
Recommended calibration 1 year
period
15-6