EasyManuals Logo

YOKOGAWA GS200 User Manual

YOKOGAWA GS200
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #179 background imageLoading...
Page #179 background image
15.6 General Specifications
Item Specifications
Display
256 × 64 dots fluorescent tube (VFD)
Internal memory GS200ROM: 4 MB (non-volatile; used to save setup files and program files)
GS200RAM: 16 MB (volatile, cleared when the power is turned off; used to save the measured
results)
Warm-up time
Approx. 60 minutes
Storage conditions Temperature –15°C to 60°C
Humidity 20% RH to 80% RH (no condensation)
Elevation 2000 m or less
Operating conditions Temperature 5°C to 40°C
Humidity 20% RH to 80% RH (no condensation)
Elevation 2000 m or less
Rated supply voltage 100 VAC, 120 VAC, and 230 VAC
Permitted supply voltage Each rated voltage ± 10%
range
Rated power supply 50/60 Hz
frequency
Permitted power supply 48 Hz to 63 Hz
frequency range
Maximum allowable input
Across OUTPUT Hi and OUTPUT Lo and across SENSE Hi and SENSE Lo ±32 Vpeak
voltage Across OUTPUT Lo and G terminal and across SENSE Lo and G terminal
±42 Vpeak
Across each OUTPUT terminal and each SENSE terminal ±0.5 Vpeak
Across G terminal and case ±250 Vpeak
Maximum power Approx. 80 VA
consumption
Maximum common-mode Across each terminal and case
±250 Vpeak
voltage
External dimensions Approx. 213 (W) × 88 (H) × 379 (D) mm (excluding protrusions)
W
eight Approx. 5 kg
Recommended calibration 1 year
period
15-6
IM GS210-01EN

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA GS200 and is the answer not in the manual?

YOKOGAWA GS200 Specifications

General IconGeneral
BrandYOKOGAWA
ModelGS200
CategoryPower Supply
LanguageEnglish

Related product manuals