Appendix 1 Factory Default Settings
Item Factory Default
Setting
Saved When
the GS200
Turns Off?
Saved to the
Setup Data File?
Source
Source function Voltage No Yes
Source range 1 V No Yes
Source level 0 V No Yes
Voltage limiter 30 V No Yes
Current limiter 200 mA No Yes
Output on/off Off No No
Program file Unselected
*
No Yes
Program repetition mode On No Yes
Program interval time 0.1 s No Yes
Program slope time 0.0 s No Yes
Program trigger Norm No Yes
Measurement (Option)
Measurement on/off On No Yes
Measurement trigger Timer No Yes
Measurement timer 0.1 s No Yes
Integration time 1 PLC No Yes
Measurement delay time 0 ms No Yes
NULL computation on/off Off No Yes
Storage on/off Off No No
Storage count 10000 No Yes
Storage file creation On No Yes
Wiring system 2W No Yes
Guard terminal Off No Yes
External I/O BNC IN terminal TRIG No Yes
BNC OUT terminal TRIG No Yes
Communication
USB feature selection Storage
*
Yes No
GP-IB address 1
*
Yes No
GP-IB command mode GS200 mode
*
Yes No
DHCP on/off On
*
Yes No
IP address 0.0.0.0
*
Yes No
Subnet mask address 0.0.0.0
*
Yes No
Default gateway address 0.0.0.0
*
Yes No
Transmission terminator CR+LF
*
Yes No
Other settings
File that is loaded when the
GS200 turns on
Unspecified
*
Yes No
Display brightness 4
*
Yes No
Display on/off On
*
No No
Beep sound on/off On
*
Yes No
Date and time Japan Standard
Time
*
Yes No
Time difference from GMT
(Greenwich Mean Time)
+9:00
*
Yes No
CSV file format Decimal point: .
*
Yes No
Delimiter: ,
*
Yes No
* This setting is not initialized to its factory default setting when you initialize setup data (see
section 9.1) or send the
*RST
command.
App-1
IM GS210-01EN
Appendix
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
App
Index
Appendix