4 - 10 MP7000 Scanner Scale Integrator Guide
10.00 Position 2 10.00 .01 lb .005 lb
10.00 Position 3 10.00 .01 lb .005 lb
10.00 Position 4 10.00 .01 lb .005 lb
Table 4-5 Verification Tests for Metric Scale
Verification Test for Metric
Scale Applied Load kg
Indication = Applied Load
Within Applicable
Tolerances
All Tolerances Are
+/-
All Tolerances Are
+/-
Acceptance Single
Int
Acceptance Dual
Int
Increasing Load Test
0.000 0.000 0.0 kg 0.0 kg
0.100 0.100 .0025 kg .001 kg
2.500 2.500 .0025 kg .002 kg
5.000 5.000 .005 kg .003 kg
10.000 10.000 .005 kg .005 kg
15.000 15.000 .0075 kg .0075 kg
Over Capacity Test
Over Capacity 15.100 EEEE N/A N/A
Decreasing Load Test
15.000 15.000 .0075 kg .0075 kg
10.000 10.000 .005 kg .005 kg
5.000 5.000 .005 kg .003 kg
2.500 2.500 .0025 kg .002 kg
0.100 0.100 .0025 kg .001 kg
Return to Zero Test
0.000 0.000 0.0 kg 0.0 kg
Shift Test (see
Figure 4-5)
5.00 Position 1 5.000 .005 kg .003 kg
Table 4-4 Verification Tests for US (lb) Scales (Continued)
Verification Test for US Scale
Applied Load lb
Indication = Applied Load
Within Applicable
Tolerances
All Tolerances Are
+/-
All Tolerances Are
+/-
Acceptance Single
Int
Acceptance Dual
Int