4 - 12 MP7000 Scanner Scale Integrator Guide
Audit Tallies
Audit Tallies (calibration counter = C, legal parameter counter = P, Legally Relevant firmware version number = F) 
used for scale verification can be shown on the Scale Display, or the 7-segment display.
1. Press and hold the Scale Zero button for three seconds to access the calibration tallies (not within Calibration 
mode). 
a. C###, P###, and #.##F tallies blink on the Scale Display and/or scroll on the 7-segment display. 
i. C### represents the number of times the scale reached 
Calibration Success on page 4-7. (Only 
successful calibrations are counted.) 
ii. The P### represents the number of times any sealable parameter, or legal parameter was changed 
with or without reaching 
Calibration Success on page 4-7. #.##F represents Legally Relevant firmware 
version number.
2. When the Scale Zero button is released, the scale returns to normal operation, and the tallies are removed 
from both displays.
3. After verification, seal the scale by recording/reporting Audit Trail or apply physical seal to models that contain 
a physical calibration switch. Report information as required by local Weight & Measures Regulations, or 
where required by law.
NOTE This feature is inhibited when the unit is in a scale Calibration Mode.
If units of weight measure was modified (i.e., kg to lb, or lb to kg) the parameter counter is incremented by two in 
the audit tally. This is because the Automatic Zero Setting legal parameter is being enabled when going from lb 
to kg, and disabled when going from kg to lb.