Scale Calibration (For Models With a Scale) 4 - 5
Legal Scale Dampening Filter
Set the scale sensitivity to vibration by scanning the appropriate barcode (see page 
4-18). To program this 
parameter, the scale must be in Calibration Mode. Barcode command values are as follows:
• 0 = Higher (most sensitive to vibration)
• 1 = Low (factory default) 
• 2 = Very Low
• 3 = Ultra Low
The lower the number, the more sensitive the scale is to vibration. The higher the number, the slower the scale may 
operate.
Step 4 - Calibration at NO LOAD
When Calibration Mode enters properly, the Scale Display blinks between CAL00 and CAL _ _ and the 7-segment 
display scrolls C00Lb, or C00g. Scale calibration can begin.
1. Ensure the platter is installed and there is NO WEIGHTED LOAD on it.
2. Touch the Scale Zero button on the front panel.
3. If Calibration at NO WEIGHTED LOAD is successful after the Scale Zero button is pressed, the process 
continues with 
Step 5 - Calibration at LOAD on page 4-5.
or
If Calibration at NO WEIGHTED LOAD is not successful after the Scale Zero button is pressed, the process 
continues with 
Calibration Failure on page 4-8.
Step 5 - Calibration at LOAD
• The Scale Display blinks either CAL25 or CAL11, depending on the units of measure programmed (CAL25 = 
pounds; CAL11 = kilograms).
• The diagnostic 7 -segment display scrolls C25Lb or C11g.
IMPORTANT This step of the calibration procedure must be completed within 30 seconds. If this time expires before 
completing the step then the calibration procedure is exited.
IMPORTANTThis step of the calibration procedure must be completed within 30 seconds. If this time expires before 
completing the step then the calibration procedure is exited.