MODULE 4: RUN OPTIONS
4-22 AEROSET
®
Customer Training Guide
94858-103 — July 2003
• If the System status is SAMPLING END, perform the following
steps:
1. Order the patient samples in the Order Samples
screen.
2. Confirm that the red Rerun Indicator Light is not
illuminated and that the System status is SAMPLING
END before performing this procedure.
3. Select <PAUSE> from the Action Area. The RUN
OPTIONS screen is displayed.
4. Select <Pause> in the right-hand column of the RUN
OPTIONS screen to pause sampling.
5. Wait for the System status to change to PAUSE.
6. Place the sample in the specified carrier and position.
7. Place the Sample Carrier on a free section of the
Sample Carrier Tray.
8. Select <UNPAUSE> in the Action Area. The RUN
OPTIONS screen is displayed.
NOTE: Do not select <RUN> to reinitiate sampling. If
<RUN> is selected, all samples remaining on the System
with tests still in process are sampled again.
9. Ensure that the Patient option is still selected.
10. Select <Unpause> in the right-hand column of the
RUN OPTIONS screen to resume sampling.
NOTE: The Operator can select <UnpsTrk>, if desired,
and the Sample Carousel remains in the PAUSE status.