Activity
AEROSET
®
Customer Training Guide 6-33
94858-103 — July 2003
ACTIVITY
Introduction
In this exercise, you will order calibrations, load the samples, and review
the calibration curves upon completion.
Objectives
Upon completion of the Calibration Activity, you will be familiar with:
Ordering a full calibration
Ordering QC to run automatically after calibration
Loading calibrators and controls
Reviewing and printing calibration data
Deleting calibration curves
Necessary Materials
Reagents:
Albumin, ALT, AST, Calcium, Creatinine, Glucose, Phosphorus,
Total Protein, Urea, and Iron
Other:
Abbott Multiconstituent Calibrators
Abbott Iron/Magnesium Calibrators
Quality Control Material
Procedure
1. Select <RUN> in the Action Area.
2. Place a checkmark (;) in the box to the left of the Calibration
option by touching the box.
3. Place a checkmark (;) in the box to the left of the QC after Calib
option by touching the box.
4. Select the Checkbox under the desired QC levels.
Instructor will need to get Iron/Mg
Calibrators, MCC’s, and QC
material from the refrigerator.
Depending on when activity is done,
may also need ICT Calibrators.