Activity
AEROSET
®
Customer Training Guide 9-11
94858-103 — July 2003
ACTIVITY
Introduction
In this activity, you will remove and replace the ICT™ Unit components
and verify the operation of the ICT Unit by visually inspecting the
components and by performing calibration.
Objectives
Upon completion of the ICT Module Activity, you will be familiar with:
• Removing and replacing the ICT Module
• Removing and replacing the ICT Probe
• Aligning the ICT Probe
• Ordering an ICT calibration and loading the calibrators and controls
• Reviewing the ICT calibration information in the ASSAY STATUS
and CALIBRATION SUMMARY screens
Necessary Materials
Reagents:
• ICT Sample Diluent
Other:
• ICT Serum Calibrators
• Quality Control Materials