Error Codes and Messages
AEROSET
®
Customer Training Guide 11-5
94858-103 — July 2003
ERROR CODES AND MESSAGES
Result Flags
Result Flags are displayed for patient and control results on the screens
indicated in the table below. They can also be printed on the patient and
control reports.
Result Error Codes
Result Error Codes are displayed on the RESULT and Reaction Graph
screens and can be printed on the Patient Reports.
• The Result Error Codes are assigned a priority with 1 = highest
priority and 14 = lowest priority. The priority does not display on
the SCC computer.
• If more than one Result Error Code is generated for a result, the
Result Error Code with the highest priority is displayed in the
RESULT screen and printed on the Patient Report.
• If more than one Result Error Code of the same
priority is generated
for a result, the first error to occur is displayed in the RESULT
screen and printed on the Patient Report.
• To view all Result Error Codes generated for a result, access the
Reaction Graph screen.
• If a “C” appears with a result on the DATABASE screen, this
indicates that a Result Error Code has been generated for the
sample.
Type of
Result
Screen “L” “H”
Patient DATABASE and
RESULT
Indicates that the
result is less than
the defined
Reference-L
Indicates that the
result is greater
than the defined
Reference-H
Quality
Control
DATABASE,
RESULT,
ASSAY STATUS,
Levey-Jennings
Graph, QC Details,
and QC SUMMARY
Indicates that the
QC result is greater
than
–2 SD from the
mean
Indicates that the
QC result is greater
than
+2 SD from the
mean
• Ensure Ops Manuals are
available for use with the Error
Codes and Messages section.
• On database, will see “C” for
error code.
• Result error code and result, if
present will be displayed in
Result screen.