MODULE 6: CALIBRATION
6-38 AEROSET
®
Customer Training Guide
94858-103 — July 2003
ANSWERS
1. a. Factor
b. Multi-point linear
c. Spline
d. Use Factor/BLK
e. 1-Point Linear
f. Abs
2. Full
3. No, it does not require
calibration, but the Operator may choose to
calibrate the assay, if necessary, to resolve the concern.
4. There is a green line below the calibrator name on the button.
5. The ASSAY STATUS screen
6. If the factors from the new and the previous curve vary by more than
the defined FAC limit %.
7. a. CALIBRATION SUMMARY (stored)
b. Calibration Details
8. The assay button text in the ASSAYS screen is yellow, a Calibration
Error Code is displayed in the ASSAY STATUS screen, the CAL
Error Code is displayed in the RESULT screen, the calibration
information in the CALIBRATION SUMMARY screen is
displayed in brown, and a calibration error code is displayed in the
Error Log.
9. The Set 1, Set 2, or Set 3 option must be selected in the RUN
OPTIONS screen.
10. No