Calibrating an Assay
AEROSET
®
Customer Training Guide 6-13
94858-103 — July 2003
CALIBRATING AN ASSAY
Ordering Calibration
To order a calibration for one or more assays, perform the following
procedure:
1. Select <RUN> in the Action Area. The RUN OPTIONS screen is
displayed.
Figure 6.1: RUN OPTIONS Screen
2. Place a checkmark (;
;;
;) in the box to the left of the Calibration
option by touching the box.
3. Touch the Drop-down list box next to Calibration and select the
Calibrator/Control Carousel and Sample Carousel
configurations desired.
Option Calibrator Locations
The System only uses calibrators on the Calibrator/Control Carousel.
Set 1 The System uses calibrators on the Calibrator/Control Carousel and the
Set 1 configuration of the Sample Carousel.
Set 2 The System uses calibrators on the Calibrator/Control Carousel and the
Set 2 configuration of the Sample Carousel.
Set 3 The System uses calibrators on the Calibrator/Control Carousel and the
Set 3 configuration of the Sample Carousel.