EasyManua.ls Logo

Abbott Diagnostics Aeroset - Page 218

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODULE 6: CALIBRATION
6-14 AEROSET
®
Customer Training Guide
94858-103 — July 2003
4. Select the return arrow.
NOTE: The System does not allow the Sample Carousel
to be used for calibrators and patient samples at the same
time. If the Set 1, Set 2, or Set 3 options are selected, the
Operator can only place a checkmark (;) by either the
Patient or Calibration option. The other option is
automatically deselected.
Example: The RUN OPTIONS screen has a checkmark
(;) by the Calibration option and Set 1 configuration is
selected. If the Operator selects the Patient option, the Cal-
ibration option is no longer selected.
If the blank (_______) Calibration option is selected, then
only patient samples can be loaded in the Sample Carou-
sel and calibrators can only be loaded in the Calibrator/
Control Carousel. The System then allows both the Cali-
bration and the Patient options to be selected.
5. Select <Order> to the right of the Calibration option. The
SELECT ASSAYS FOR CALIBRATION screen is displayed.
Figure 6.2: SELECT ASSAYS FOR CALIBRATION Screen
6. Select the full calibration option.
7. Select the assay button for the assays to be calibrated. Individual
assays or panels may be selected.
The assays last selected and saved
are highlighted. Deselect, then
select new ones.

Table of Contents