MODULE 10: MAINTENANCE
10-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
ANSWERS
1. 1% Alkaline Wash Solution
2. Alcohol
3. Bubbles in the tubing or drips from the ICT™ Probe
4. To avoid dripping water from the bottom the Reagent Supply
Center segments into the open Reagent Cartridges
5. To remove spilled or splashed control or sample material to
minimize the biohazard
6. R2-B
7. a. New assay installation and calibration
b. Values for a new lot of multiconstituent calibrators are entered
c. QC means are adjusted or edited for a new lot of controls
d. Changes are made to System configuration files
e. Robotic adjustments are made by the Customer or Field Service
Representative
f. Customer determined interval for System backup to capture QC
or calibration data
8. A formatted floppy disk
9. a. Ensure that the bottom of the syringe barrel is in the groove in
the syringe holder.
b. Ensure that the plunger flange is below the U-shaped bracket.
10. Quarterly
11. c. spacer
12. Check the inlet and outlet ports with water. Ensure that water can
not be drawn in through the outlet port, and water can not be
expelled through the inlet port.