Calibration Options
AEROSET
®
Customer Training Guide 6-11
94858-103 — July 2003
CALIBRATION OPTIONS
In addition to full calibration, which renews all data points, three other
options of calibration correction are available for photometric tests. After
correcting the absorbance data, a new calibration curve is generated.
Listed below are the four calibration types.
• Full Calibration
• Blank Calibration
• 1-Point Adjustment
• 2-Point Adjustment
Full Calibration
All data points (reagent blank and calibrators) specified on the
Calibration page of the Assay Configuration screen are reanalyzed and
a new calibration curve is generated.
NOTE: A full calibration is required to update the calibration
date and time.
Other Calibrations
See the AEROSET System Operations Manual for additional
information about Blank Calibrations, 1-Point Adjustments, and 2-Point
Adjustments.