Calibration Review
AEROSET
®
Customer Training Guide 6-19
94858-103 — July 2003
CALIBRATION REVIEW
After calibration is performed, the AEROSET System verifies the
calibrator results by comparing them to the assay parameter
specifications defined on the Calibration page of the Assay
Configuration screen.
The AEROSET System software requires one valid calibrator replicate
per required level to compute a curve, unless a greater number is
specified in the assay-specific Package Inserts.
NOTE: Refer to the System Parameters section of the assay-
specific Package Inserts for information on the minimum
number of replicates for a specific assay.
If no System errors occur and the calibration results are within the assay
parameter specifications, the calibration passes. The assay button text in
the ASSAYS screen is green.
If a System error occurs or if the calibration results fall outside the assay
parameter specifications the following occurs:
The assay button text in the ASSAYS screen is yellow.
A Calibration Error Code is displayed in the ASSAY STATUS
screen and Error Log.
The Result Error Code CAL is displayed in the RESULT screen for
any patient sample run after the failed calibration.
Go to Activity and complete the
Procedures and Loading the
Calibrators portions. Return to lecture
with Calibration Review.
Remind to check QC after cal in Run
Options.
Uses actual absorbance from 1 rep,
average absorbance from 2 reps; for 4
or 5 reps it will throw out low and
high, and average remaining.