EasyManua.ls Logo

Abbott Diagnostics Aeroset - Page 222

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODULE 6: CALIBRATION
6-18 AEROSET
®
Customer Training Guide
94858-103 — July 2003
System-Initiated Calibration
When the AEROSET System determines that a calibration is required at
the time a patient run is initiated and the Calibration option is not
selected on the RUN OPTIONS screen, the System displays a screen to
inform the Operator of this condition. The Operator must determine if
the calibration is required, based on anticipated or current samples to be
processed.
Procedure for System-Initiated Calibration
1. Select <Start> in the RUN OPTIONS screen to initiate a run. The
System determines that a calibration is required, and a screen is
displayed indicating which calibrations are required.
2. Select <OK>. The SELECT ASSAYS FOR CALIBRATION
screen is displayed. Four options are available.
3. Select the desired option:
4. The Operator must load any necessary calibrators.
Option Description
<OK> All suggested calibrations and automatic
calibrations are performed.
<Cancel> No displayed calibrations or automatic
calibrations are performed.
NOTE: Selecting <Cancel> does NOT
prevent the AEROSET System from
running. <Cancel> refers only to the
Calibration option being displayed.
<DSelect> All assays are deselected. Only automatic
calibrations are performed.
NOTE: After selecting <DSelect> the
Operator must then select <OK>.
Manual Selection Select or deselect the desired assay(s). All selected
calibrations and automatic calibrations are
performed.

Table of Contents