Calibrating an Assay
AEROSET
®
Customer Training Guide 6-17
94858-103 — July 2003
4. If the Set 1, Set 2, or Set 3 options are selected on the RUN
OPTIONS screen, select the corresponding button on the
CALIBRATOR/CONTROL screen to view the calibrators that
need to be loaded.
NOTE: I-Index (index solution) is not currently used, but
Calibrator/Control position 42 (I-Index) must have a
concentration of zero defined for all ICT™ assays. If the
concentrations are not defined, the ICT buttons do not
display on the SELECT ASSAYS FOR CALIBRATION
screen.
Initiating the Calibration Run
1. Select <RUN> from the Action Area of the Main Display to display
the RUN OPTIONS screen.
2. Confirm that the Calibration option is still selected.
3. Select <Start> to initiate the run.