EasyManua.ls Logo

Abbott Diagnostics Aeroset - Loading the Calibrators

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MODULE 6: CALIBRATION
6-16 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Loading the Calibrators
1. Select <Cal/Control> in the Information Access Area. The
CALIBRATOR/CONTROL screen is displayed. The calibrators
required to perform the selected assays are highlighted with a green
line under the Calibrator Name on each button.
Figure 6.4: CALIBRATOR/CONTROL Screen
NOTE: If the QC after Calib option has been selected, the
required QC samples are highlighted with a yellow line
under the QC name on each button.
2. Pipette the appropriate calibrator materials into sample cups.
NOTE: When water is defined for the blank, water is
dispensed by the System from the sample probe.
Therefore, water does not need to be loaded on the
Calibrator/Control Carousel.
3. Load the calibrators in the carousel positions defined on the
CALIBRATOR/CONTROL screen.
Show MCC’s
A lot change in calibrators will have
a corresponding change in values.
Point out the need to use AEROSET
values.
MCC Calibrators expire 7 days after
opening.

Table of Contents