Solutions Used for START UP, SHUTDOWN, and
AEROSET
®
Customer Training Guide 3-15
94858-103 — July 2003
SOLUTIONS USED FOR START UP, SHUTDOWN, AND
DAILY OPERATION
The following table describes the solutions used for each START UP
and SHUTDOWN option and for Daily Operations. It also indicates the
placement of the solutions.
Solutions for START UP and SHUTDOWN
Option Solution Used Placement
Change Water In Bath Water Bath Additive
a
Reagent Supply Center 1, outer segment
Wash Cuvettes Alkaline Wash Solution On the weight platform behind the front-right door
Acid Wash Solution On the weight platform behind the front-right door
Fill with Dtg Optional detergent
(for future application)
• Reagent Supply Center 1, outer segment
• Reagent Supply Center 1, inner segment
Wash Probes with Water None
1% Alkaline Wash
Solution
b
• Calibrator/Control Carousel, Position 44 (1 mL)
• Reagent Supply Center 1, outer segment
• Reagent Supply Center 1, inner segment
• Reagent Supply Center 2, outer segment
• Reagent Supply Center 2, inner segment
Wash ICT™ with ICT Cleaning Fluid
d
Calibrator/Control Carousel Position 43 (200 µL)
ICT Sample Diluent Reagent Supply Center 1, outer segment
ICT Reference Solution On the weight platform behind the front-right door
Drain & Fill IRef Cup ICT Reference Solution On the weight platform behind the front-right door
Flush Water Lines None None
Solutions for Daily Operation
Option Solution Used Placement
Wash Cuvettes Alkaline Wash Solution On the weight platform behind the front-right door
Acid Wash Solution On the weight platform behind the front-right door
SmartWash™ 1% Alkaline Wash
Solution
b
• Calibrator/Control Carousel, Position 44
• Reagent Supply Center 1, outer segment
• Reagent Supply Center 1, inner segment
• Reagent Supply Center 2, outer segment
• Reagent Supply Center 2, inner segment
0.5% Acid Wash
Solution
c
• Calibrator/Control Carousel, Position 45
• Reagent Supply Center 1, outer segment
• Reagent Supply Center 1, inner segment
• Reagent Supply Center 2, outer segment
• Reagent Supply Center 2, inner segment
ICT Analysis ICT Sample Diluent Reagent Supply Center 1, outer segment
ICT Reference Solution On the weight platform behind the front-right door
Sample Dilution Saline • Reagent Supply Center 1, outer segment
• Reagent Supply Center 1, inner segment