MODULE 3: DAILY START UP AND CHECKS
3-14 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Automatic Power ON, START UP, and SHUTDOWN
The AEROSET System can be configured to Power ON, perform
START UP and SHUTDOWN options automatically, even when the
Operator is not present.
On the System page of the MAINTENANCE UTILITIES screen, the
time and mode (Power ON, START UP, SHUTDOWN) are defined for
each day of the week.
The Rotary Power Control Switch must be set in the AUTO position
for the preset option to execute at the specified time.
To configure the Automatic Power ON, START UP, and SHUTDOWN
options, select the Maintenance Utilities icon in the Action
Area to access the MAINTENANCE UTILITIES screen.
For information about configuration of the Automatic Power ON,
START UP, and SHUTDOWN options, refer to MAINTENANCE
UTILITIES Screen in Section 2, Installation Procedures and Special
Requirements of the AEROSET System Operations Manual.
The System must first be powered OFF and the Rotary Power
Control Switch set to the AUTO position for the automatic
Power ON, START UP, and SHUTDOWN options to engage.
Only one option/day.
If you are running the
instrument at the time of auto
START UP or SHUTDOWN, it
won’t do them.