MODULE 6: CALIBRATION
6-10 AEROSET
®
Customer Training Guide
94858-103 — July 2003
CALIBRATION FREQUENCY
Required Calibration
You must perform a full calibration run when:
• The Abbott AEROSET System is installed.
• The calibration curve has expired.
• A new reagent lot number is used.
• Existing assay file parameters have been updated or changed.
• A new assay file is installed.
• An Abbott Field Service Representative replaces the System
Control Center (SCC) and assay files have not been backed up. It is
recommended that a backup be performed routinely.
• It is required per assay application.
• An ICT™ component is removed and replaced.
• The ICT Reference Solution is replaced.
Optional Calibration
You may need to perform a new calibration run on an assay in the
following situations:
• Assay control values are out of specification.
• Any System component is replaced.
• Any System maintenance procedure is performed.
• Assay-specific calibration checks are not within specification.