MODULE 6: CALIBRATION
6-10 AEROSET
®
Customer Training Guide
94858-103 — July 2003
CALIBRATION FREQUENCY
Required Calibration
You must perform a full calibration run when:
The Abbott AEROSET System is installed.
The calibration curve has expired.
A new reagent lot number is used.
Existing assay file parameters have been updated or changed.
A new assay file is installed.
An Abbott Field Service Representative replaces the System
Control Center (SCC) and assay files have not been backed up. It is
recommended that a backup be performed routinely.
It is required per assay application.
An ICT™ component is removed and replaced.
The ICT Reference Solution is replaced.
Optional Calibration
You may need to perform a new calibration run on an assay in the
following situations:
Assay control values are out of specification.
Any System component is replaced.
Any System maintenance procedure is performed.
Assay-specific calibration checks are not within specification.