MODULE 11: COMPONENT REPLACEMENT AND TROUBLESHOOTING
11-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
ANSWERS
1. The Reaction Graph screen
2. “bye”
3. An Error Log Message is generated and the cuvette position is
displayed with a red background on the Cuvette Integrity screen in
MAINTENANCE UTILITIES.
4. The lamp housing cover was not correctly seated.
5. Any six of the following:
a. Setting the date and time
b. Defining the Automatic Power ON, START UP, and
SHUTDOWN
c. Component tracking
d. Assay counts
e. System Software Backup and Restore
f. Formatting a floppy disk
g. Library Maintenance
h. Manually moving robotics
i. Scan bar codes
j. View cuvette integrity
k. View electronic A/D data for the photometer and ICT™
Unit
l. View the current software version
6. Poppet Valves
7. 61mm