MODULE 7: QUALITY CONTROL
7-12 AEROSET
®
Customer Training Guide
94858-103 — July 2003
REVIEWING QC RESULTS
When QC results are completed with no errors, the results are saved and
the QC statistics are adjusted accordingly. The QC values and statistics
can be viewed in the ASSAY STATUS, Levey-Jennings Graph, QC
Data List, and QC SUMMARY screens. The data can be printed on the
Levey-Jennings Graph or QC Data List forms. The QC data can also be
exported to a floppy disk.
Expected concentration and SD can be defined for the QC, and Westgard
Rules can be selected to evaluate the QC. If a Westgard Rule is violated,
an Error Log Message is generated, and the rule displays next to the QC
data in the ASSAY STATUS screen.
NOTE: Refer to Section 2, Installation Procedures and Special
Requirements of the AEROSET System Operations Manual
for details on QC configuration.