MODULE 8: CONFIGURATION
8-26 AEROSET
®
Customer Training Guide
94858-103 — July 2003
ASSAY CONFIGURATION
Assay configuration includes the following procedures:
• Reagent Configuration
• Calibrator Configuration
• Control Configuration
• Assay Parameter Configuration
Reagent Configuration
No reagent configuration is required for bar code labeled reagent
cartridges supplied by Abbott Laboratories. Select <RGT SCAN> from
the Action Area of the Main Display after the reagent cartridges have
been loaded.
For reagent cartridges without bar code labels, configuration is
performed by selecting <Reagents>, and then selecting the button for the
position in which the reagent will be loaded. The following information
is defined:
• Steps for multiple Assay
Configuration (can also be
used for single):
1.Load reagents and scan bar codes.
2. Configure Cals and Controls
names.
3. Enter calibrator values and
control means in the Cal and
Control screen.
4. Configure all pages of Assay file.