Assay Configuration
AEROSET
®
Customer Training Guide 8-27
94858-103 — July 2003
For information about configuration of reagents, refer to Reagent
Configuration in Section 2, Installation Procedures and Special
Requirements of the AEROSET Systems Operations Manual.
Field Description
Reagent Name Allows entry of up to 7 characters. This is a required field.
CAUTION: Verify the entered reagent name is identical
to the reagent name on the Base and SmartWash™
pages of the Assay Configuration screen. If the reagent
name is not identical on the Base and SmartWash pages,
SmartWash will not occur and erroneous results could be
produced.
Reagent Type This is a required field. One of the following is selected:
R1—Reagent 1
R2—Reagent 2
Diluent—sample diluent
AlkW—onboard diluted alkaline wash
AcidW—onboard diluted acid wash
Dtg—optional detergent as a wash solution
Additive—water bath additive
Cartridge Size The size of the cartridge is selected to allow for correct
estimation of the remaining tests. This is a required field.
Exp Date Reagent expiration.
Enter the reagent expiration date. When the expiration date
is reached and a reagent scan is complete, an Error Log
Message displays.
NOTE: The Exp Date field must be configured to
generate an expired (EXP) Result Error Code and
Error Log Message.
Lot # Reagent lot number.
Cartridge # Cartridge serial number, if available.
Rgt (Reagent)
Lo Alert
An alert will occur when the reagent cartridge volume is
less than the %volume or number of tests defined in this
field.
NOTE: If this option is used, it overrides the alert
configured through System configuration.
• First three fields are required.
!