System Configuration Activity
AEROSET
®
Customer Training Guide 8-25
94858-103 — July 2003
Procedure
1. Create the following “super” level User Code/Password:
• big / tex
2. On the Printer A Configuration screen make the following edit:
• Deselect the Auto-Print Patient Reports option
3. On the Printer B Configuration screen make the following edits:
• Result Printing Mode > All
• Deselect the Auto-Print Control Reports option
• Select the Auto-Print Patient Reports option
4. Edit the following on-line communication options to:
• Protocol Type > Type 1’
• Result Transmission Mode > All
• Deselect the Transmit QC Results option
• Deselect the Use Transmission Check Digit option
• RS-232 Config > 8L-1S-NP and 9600
5. On the PAGE SETUP screen add IgG assay to Form 2. Place the
assay below the “pack” line in the first available space.
6. Create the following panel:
• Panel Name - Day QC
• Panel Number - 250
• Assays - Na, K, Cl, Glu, Ca, AlbP, TP
7. Load all the bar coded tubes into carriers and initiate the run by
selecting <RUN> in the Action Area. Ensure that only the Patient
option is selected and select <Start>.
• Point out that printer
configuration steps in the
activity change default form
from Form 1 to Form 2.