Ordering Controls
AEROSET
®
Customer Training Guide 7-11
94858-103 — July 2003
When the Carousel Sampler has Completed Sampling
1. Confirm that the Sample Carousel Movement Indicator Light is not
illuminated or blinking.
2. Place the controls (in sample cups) in the Calibrator/Control
Carousel in the specified positions.
3. Select <PAUSE> in the Action Area. The RUN OPTIONS screen
is displayed.
4. Select <Pause> in the right-hand column of the RUN OPTIONS
screen to pause sampling.
5. Wait for the System status to change to PAUSE.
6. Select <UNPAUSE> in the Action Area. The RUN OPTIONS
screen is displayed.
NOTE: Do not select <RUN> to reinitiate sampling. If
<RUN> is selected, all samples remaining on the System
with tests still in process, are sampled again.
7. Ensure that the desired control option is selected in the RUN
OPTIONS screen.
NOTE: Start QC is not performed if the run is restarted
using the PAUSE and UNPAUSE commands.
8. Select <Unpause> in the right-hand column of the RUN
OPTIONS screen to resume sampling.
NOTE: The Operator can select <UnpsCrsl> which
unpauses the carousel only.