MODULE 6: CALIBRATION
6-34 AEROSET
®
Customer Training Guide
94858-103 — July 2003
5. Touch the Drop-down list box next to Calibration and select the Set
1 option.
6. Select the return arrow to close the list.
7. Select <Order> to the right of the Calibration option.
8. Ensure that the calibration type of Full is selected.
9. Select the following assays: AlbP, ALT, AST, Glu, Iron, TP, Phos,
Crea, Urea, and CA.
10. When all desired assays have been selected, select <OK>.
11. Select <Save> in the right-hand column.
12. Select <OK> on the confirmation dialog window.
13. Select <Cancel> on the RUN OPTIONS screen to exit the screen
without starting the run.
Loading the Calibrators
1. Select <Cal/Control> in the Information Access Area.
2. Load the highlighted calibrators and controls in the indicated
positions on the Calibrator/Control Carousel.
3. Select <Set 1>.
4. Load the highlighted calibrators in the indicated positions on the
Sample Carousel.
5. Select <OK> to exit the CALIBRATOR/CONTROL screen.
6. Select <RUN> in the Action Area.
7. Confirm that the Calibration (with Set 1) and QC after Calib options
are still selected.
NOTE: Ensure that the Patient option is not selected.
8. Select <Start> to initiate the run.