MODULE 4: RUN OPTIONS
4-30 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Add a Sample in the Reserved STAT Position
1. Confirm that the STAT Sampling Indicator Light is not illuminated
and place the sample in the Reserved STAT Position.
2. Select <STAT> from the Action Area.
3. Enter the following information:
4. Select <Start>.
Add Samples During a Run
1. Select <Database> from the Information Access Area.
2. Select <Order>.
3. Select <New>, if necessary, to display the Order Samples screen.
4. Order the following patient samples in the Sample Carousel:
5. Select <PAUSE> in the Action Area.
6. Select <Pause> in the right-hand column of the RUN OPTIONS
screen.
7. Wait for the System status to change to PAUSE and then place the
samples in the Sample Carousel in Section B, Positions 1 - 3.
8. Select <UNPAUSE> in the Action Area.
9. Confirm that the checkmark (;
) in the box to the left of the Patient
option is still present. Confirm that the Car’r number defined for
Sample Carousel, Section B is 2000.
10. Select <Unpause>.
SID Assays
789STAT Chem1
SID Name C/P Assays
3050 Tannic, Ty 2000 / 1 Chem2, Iron
3067 Ester, Polly 2000 / 2 ICT, Phos, AST, ALT
3071 Kew, Barbie 2000 / 3 Glu, TP, AlbP, Iron