Westgard QC Rules
AEROSET
®
Customer Training Guide 7-41
94858-103 — July 2003
NOTE: QC data are checked first against the 1-2 SD rule, if
enabled. If the data are within 2 SD, the data are considered
“in-control” and are not evaluated against the other QC rules. No
Westgard rule Error Log Message appears on the ASSAY
STATUS screen when the data are considered “in-control”.
An example of a Westgard Rule violation is shown below: