MODULE 8: CONFIGURATION
8-36 AEROSET
®
Customer Training Guide
94858-103 — July 2003
NOTE: The maximum volume of sample, reagent, and as
needed water must not exceeded 360 uL.
NOTE: Total volume of sample, reagent, and water must be at
least 160 uL to ensure proper mixing and provide adequate
volume for photometry.
R.Vol The reagent volume is defined for Reagent 1 and Reagent 2 (if required).
Range is 10-345 µL.
NOTE: The reagent volume for ICT assays is preset to 345 µL.
W.Vol The water volume is defined when a concentrated reagent is used. Range is
0-345 µL.
Type # Controls dispense speed and over aspiration volumes. Range is 0 to 6; the
default setting is 0.
Refer to Type Number in Section 2, Installation Procedures and
Special Requirements of the AEROSET System Operations
Manual for details.
Reaction Check Used for immunologic reactions to evaluate a reaction for prozone
phenomenon.
Factor/Intercept The factor and intercept can be used to match results to another system.
NOTE: The factor/intercept is used for all assays except calculated
assays.
NOTE: If a value is entered, the quantitative result is corrected
and reported.
Decimal Places The number of decimal places to report for the result are defined. Range is
0-3.
NOTE: The decimal places are defined for all assays.
Units The result units are defined by selecting from the options in the Drop-down
list box. The Operator may also use the keyboard to enter a result unit not
available in the default list.
NOTE: Units are defined for all assays.
Field Description