MODULE 8: CONFIGURATION
8-38 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Field Description
Calib Mode The appropriate calibration curve type is selected.
Factor (for Calib Mode of
Factor only)
If the Calibration Mode is defined as Factor, the factor is entered and used for result
calculation.
Use Cal Factor from: (for Use
Factor and Use Fac/Blk only)
If the Calibration Mode is defined as Use Factor or as Use Fac/Blk, the Operator must
select the assay name of the assay calibration data that will be used.
Interval (H) The calibration interval in hours. Up to four digits can be entered.
NOTE: The calibration interval is defined for photometric and ICT™ assays.
Blank/Calib Replicates The number of replicates (up to five) for the blank and the calibrators are entered.
NOTE: The Blank/Calib Replicates are defined for photometric and ICT assays.
Extrapolation %
Defines extrapolation beyond the linearity of the assay.
NOTE: This only displays for Linear, Spline, Logit-4, Logit-5, Exponent, Use
Factor, and Use FAC/BLK Calibration Modes.
Span If the Span Data check will be performed, the Operator must define the portion of the
curve to check. The absorbance difference will be checked between the blank and
defined calibrator level.
Span Abs Range If the Span Data check will be performed, the Operator must define the acceptable
absorbance difference in the defined span.
Blank Abs Range An acceptable absorbance range may be defined for the blank.
Cal Deviation The maximum absorbance variation of the calibrator replicates may be defined.
SD For non-linear approximation curves only, the sum of the absolute values of the
difference between the calculated calibration curve, and the measured values should not
exceed the value defined in this field.
BLK The water or blank is defined in the CALIBRATOR/CONTROL screen and then
selected from a Drop-down list box.
C1-C8 The calibrator is defined in the CALIBRATOR/CONTROL screen and then selected
from a Drop-down list box.