Component Replacement
AEROSET
®
Customer Training Guide 11-21
94858-103 — July 2003
NOTE: The above table includes some of the most common
reasons for replacing the listed components and is not intended
to be a comprehensive list of all the possible reasons for
replacing components.
For information about the removal and replacement of
the sample and reagent dispense components, refer to
Component Replacement in Section 9, Service and
Maintenance of the AEROSET System Operations
Manual.
Perform the Sample and Reagent Dispense Component Replacement
Activity at the end of this Module.
Sample Carousel Tube
Bottom Holder
• Worn or damaged
• Missing or lost
Source Lamp • Photometer errors
• Lamp intensity errors
• Result imprecision
Poppet Valve • High-concentration waste not aspirating from
cuvettes
• Dripping from high-concentration waste nozzles
on the cuvette washer
• Movement not as expected
Component Reason for Replacement