GLOSSARY
Glossary-2 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Glossary
1-Point Adjustment A calibration type that uses the absorbance data for a single calibrator to adjust
the calibration curve.
2-Point Adjustment A calibration type that uses the absorbance data for the reagent blank and a
calibrator to adjust the calibration curve.
Absorbance Limit Configured range of absorbance values that are considered acceptable for
measurement purposes. Values outside this range are not used for calculation.
Absorbance Limit Check A calibration data check that evaluates the absorbance or change in absorbance
obtained during sample measurement. If the absorbance or change in
absorbance is outside the specified range a Result Error Code is generated.
Absorbance Mode A calibration mode in which results are based on the absorbance of water and
are represented as absorbance in the case of an end-point assay or as absorbance
change (rate of absorbance change per minute) in the case of a rate assay.
Accessory An item that is used repeatedly.