Biosafety Hazards
AEROSET
®
Customer Training Guide Appendix-7
94858-103 — July 2003
Decontamination Procedure Requirements
Instruments must be decontaminated prior to servicing, shipment, or
movement. For the AEROSET System, decontaminate the following
elements:
Related Information
Refer to Section 9, Service and Maintenance of the AEROSET System
Operations Manual for additional instructions for cleaning and
maintaining AEROSET System components.
Decontamination Procedures
Element Decontamination Method
Probes Avoid contact with probes. Flush the probes with a disinfectant solution, such as
0.1% sodium hypochlorite.
During maintenance procedures, clean the outside of the probes with alkaline wash
solution. Refer to Section 9, Service and Maintenance for instructions.
If probes will be reused, place in a disinfectant solution a minimum of 30 minutes
before reinstallation.
AEROSET System Surfaces and
Components
Wipe the surface or component with a detergent solution then wipe the unit with an
appropriate disinfectant, such as 0.5% sodium hypochlorite (10% solution of
chlorine bleach containing at least 5% sodium hypochlorite). Allow to air dry at
least 10 minutes.
High-concentration Waste
Container/Hose
Rinse the container with a detergent solution then rinse the container with an
appropriate disinfectant, such as 0.5% sodium hypochlorite (10% solution of
chlorine bleach containing at least 5% sodium hypochlorite).
Liquid Material Before shipment, servicing, or relocation, remove all samples, reagents, controls,
calibrators, consumable solutions, and disposables from the AEROSET System.