DD+DIS150.06E
Repair and Service
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 CR 30-X Chapter 3.5 / 63
03-2009 Type 5175 / 100/110 Agfa Company Confidential
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed (see arrow):
Figure 153
CALIBRATION:
(1) Perform a stall calibration for the roller clamping stepper motor.
Refer to chapter 3.6, Adjustments.
Verification
(1) Insert cassette with unexposed IP in the digitizer.
(2) At the workstation select examination type “System diagnosis – Flatfield”.
(3) Select “ID” to identify the image: Scanning starts.
(4) Confirm that the image arrives at the workstation.
(5) Reject the image.