DD+DIS150.06E
Repair and Service
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 CR 30-X Chapter 3.5 / 105
03-2009 Type 5175 / 100/110 Agfa Company Confidential
4.3.2 Electrical Check according to National Regulations
(1) Perform the electrical check according to National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations.
4.3.3 Verification of the successful Power Supply Replacement
Verification:
(1) Switch off/on the digitizer.
(2) Wait till the status indicator at the digitizer switches to green
(takes approx. 1,5 min.).
(3) Insert cassette with unexposed IP in the digitizer.
(4) At the workstation select examination type “System diagnosis – Flatfield”.
(5) Select “ID” to identify the image: Scanning starts.
(6) Confirm that the image arrives at the workstation.
(7) Reject the image.
Result:
The power supply is replaced.