DD+DIS147.08E
Manufacturing Standard Modifications
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0 CR 30-X Chapter 8 / 5
05-2008 Type 5175 /100/110 Agfa Company Confidential
1 Introduction of Software Patch CRE_1206C
Reason
To prevent too high displayed LgM-Values.
Product(s) &
Serial
Number(s)
Digitizer type 5175/100: SN ≥ 1170 plus 1089, 1125, 1140,
1148, 1153, 1158, 1162, 1163, 1164,
1165, 1166, 1167, 1168,
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Service Bulletin # 05, DD+DIS001.07E
2 Introduction of Software Patch CRE_1206D
Reason
To prevent interfering noise in the image.
Product(s) &
Serial
Number(s)
Digitizer type 5175/100: SN ≥ 1366 plus 1187, 1260, 1261,
1263, 1338, 1355, 1357, 1363,
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Service Bulletin # 06, DD+DIS027.07E
3 Added Post Scan Brush
Reason
Improvement of ESD (electrostatic discharge) behaviour of the
digitiser when the image plate is in the post scan area.
Product(s) &
Serial
Number(s)
Digitizer type 5175/100: SN ≥ 1048 plus 1035, 1039, 1041,
1042, 1043, 1044, 1045, 1046,
Part
Number(s)
Postscan ESD Brush Kit: CM+ 9 5175 9080 0
Referenced
Document(s)
Service Bulletin # 08, DD+DIS397.06E