142 Chapter 2
Performance Test
E5071C Performance Test Record
Port 2
System
performance
Frequency
range [Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
Directivity 100 k - 300 k < -20 0.09
Directivity 300 k -3 G < -25 0.81
Source match 100 k - 300 k < -20 0.05
Source match 300 k -3 G < -25 3.00
Load match 100 k - 300 k < -12 0.04
Load match 300 k - 3 G < -17 0.59
Transmission
Tracking
100 k - 300 k ± 1.5 ± 0.03
Transmission
Tracking
300 k - 3 G ± 1.0 ± 0.20
Reflection
Tracking
100 k - 300 k ± 1.5 ± 0.04
Reflection
Tracking
300 k - 3G ± 1.0 ± 0.04
Option 240
Port 1
Transmission
Tracking
300 k - 3 G ± 1.0 ± 0.20
Reflection
Tracking
100 k - 300 k ± 1.5 ± 0.04
Reflection
Tracking
300 k - 3G ± 1.0 ± 0.04
System
performance
Frequency
range [Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
System
performance
Frequency
range [Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
Directivity 9 k - 300 k < -20 0.09
Directivity 300 k - 3 G < -25 0.81
Directivity 3 G - 4.5 G < -20 0.40
Source match 9 k - 300 k < -20 0.05
Source match 300 k - 3 G < -25 3.00

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals