Operation using the data setting device Azbil Corporation
6-72 Model MTG11A/18A, MTG11B/18B, MTG14C
Checking pulse output [PULSE OUTPUT] function
Pulse output can be checked by fixing the pulse output from the converter main body
by using SFC.
Use the following procedure to check the pulse output.
Step
Key Procedure SFC screen
1 Confirm that the SFC is set to
“READY”.
If it is not, press the key to
set it to “READY”.
2
or
Press the key to access the
CONFIG functions. Press the
key or key to display the
screen as shown on the right
3
or
Press the key to access the
[TOTALIZER] functions. Press the
key or key to display
the screen as shown on the right
4
Press the key to display the
screen as shown on the right.
5
Press the key, again.
The actual pulse output will be
displayed as shown on the right.
6 Enter the output value to be
checked into the screen as shown
on the right using numeric key.