Operation using the data setting device Azbil Corporation
6-46 Model MTG11A/18A, MTG11B/18B, MTG14C
Changing flow rate display: [DISP] function
Introduction
It is possible to set whether an instantaneous flow rate displayed on the display panel
of the converter should be expressed as real flow rate or percentage.
Percent display refers to a percentage (%) over the maximum flow rate set by the URV.
Procedure
Use the following procedure to set or change flow rate display.
Step
Key Procedure SFC screen
1 Confirm the SFC is set to “READY”.
If it is not, press the key to set
it to “READY”.
2
Press the key to access the
CONFIG functions.
3
Press the key.
4
or
Press the key or key to
display this screen.
5 Press the [ENTER] key.
Result:
The screen as shown on the right
appears and the set flow rate
display is confirmed.
6
Pressing the key displays
DISP=%, DISP=FLOW RATE and
DISP=TOTAL by turns. Display the
screen that you want to set.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
B
CONF
SFCM00006018D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006020D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006050D
ENTER
( Yes )
NON-VOL
SFCM00006051D
MENU
I
ITEM
DE CONF.
MENU
I
ITEM
DE CONF.
SFCM00006052D
CM2-MTG300-2001.book 46 ペー   