Measuring system MAK TIGER 3003
48
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
5.19.2 Terminal connections of Bottle Drive
Bottle Drive with pneumatic lifting device type 6774-10
Inputs/ Outputs of Ultrasampler type 6771-31
25
26
27
28
29
30
31
32
33
34
+24V on-board power supply, auxiliary supply for
sensor system
Digital input IN 1
Digital input IN 2
Digital input IN 3
Digital input IN 4
Digital input IN 5
Digital input IN 6, (bottle in position)
Digital input IN 7, (star in position)
Digital input IN 8, (bottle down)
0 V on-board power supply, reference ground for
IN 1-8
As required
As required
As required
As required
As required
As required
Bottle Drive
P-Star (terminal 49)
Bottle Drive
As required
42
43
44
45
46
47
48
49
50
51
Signal M-Up, bottle lifting motor
Signal M-Down, bottle lifting motor
Signal M-Outlet, engine of outlet cassette
Signal M-Inlet, motor of inlet cassette
Signal M-star, motor of star wheel drive
Signal P-star, self-holding contact for star wheel drive
Ground reference potential for signal P-star
Signal P-star, feedback signal
+ 24 V on-board power supply line for bottle drive
GND on-board power supply line for bottle drive
Shield connection for bottle drive cable
bn
rd
ye
gn
rs
pk
gywh
wire jumper
bk
bnwh
Shield connection on-board power supply cable
+ 24 V on -board power supply feed-in
GND on-board power supply feed-in
Truck battery+
Truck battery -
+ 24 V on-board power supply output
AGND on-board power supply output
+ 44 V operating voltage output for sampler 1
rd 1,5 mm²
bu 1,5 mm²
bk 1,5 mm²
Terminal 39, sampler board
Terminal 40, sampler board
Terminal 41, sampler board
5.19.3 Terminal connections of Bottle Drive
Mini
Bottle Drive Mini type 6774-12
Outputs of Ultrasampler type 6771-31
Magnetic plate motor (via relais)
Reference ground for OUT 1-4
Reference ground for OUT 5-8